To establish connection between the device and PC, port number 6000 must be
unlocked in software and hardware firewall.
The
Export manual
button allows you to download the user manual on to the flash drive.
The
Service options
) allows the user to:
–
Delete current configuration and restore default settings (excepting Access options –
see
Chapter 7.17.
). It is done by passing the password, which is “
RENEW
CONFIGURATION
".
–
Calibrate RTC clock in the device. RTC is hardware clock which measures time
when the device is switch off. This calibration increases precision of the device and
should be made in order to compensate work temperature effect.
To do this press
Service options
button and pass the password, which is “
RTC
CALIB
” and then enter current time and date. After at least two weeks password
should be entered again and once again current time and date.
–
Calibrate system clock in the device. It is software clock which is clocked by the CPU
and it measures time while the device is switch on. During calibration a delay
between RTC and system clock is calculated and used to correct software clock
measurements. This calibration shuold be made after RTC calibration is done.
To do this press
Service options
button and pass the password, which is “
SYSCLK
CALIB
” and then the device should be powered for at least 48 hours. After this time
pass this password again.
Fig. 7.27. Device information menu
Remote display mode should not be used for normal device work, but only for its
configuration.