86
ADVIA 2400 Operator’s Guide
Exporting the sample log entries
You can export the current sample log file in ASCII format for use by another program
such as Microsoft Excel.
1.
At the File menu, select
Export
.
2.
Select or create the folder in which you want to save the file.
3.
In the File name box, type the file name you want.
The sample log file name is entered by default.
4.
In the Save as type list, select the file format.
The CSV (comma separated values) file format is selected by default.
5.
In the Deliminator list, select the character used to separate sample log data items.
6.
Select
Save
.
Reviewing the calibration results
•
Use the View Calibration Curve window to review the calibration details for
photometric methods.
•
Use the ISE Monitor to review the calibration details for electrolyte methods.
See Checking ISE Calibration for information about the criteria used to accept or reject
the ISE calibration.
Reviewing the sample results
Sample results can be reviewed on and report from the RealTime Monitor and
Review/Edit windows.
Reporting results
The system prints calibration, control, and patient sample results. Only patient and
control sample results are transmitted to a host computer.
Results are reported automatically as the sample analysis is complete and results become
available, or the operator can manually select results to report in batches.
Real-time reporting
•
During operation, the System set area of the System Monitor window manages the
printing of the sample results (usually for laboratory purposes).
•
The Automatic transfer area of the Online Settings window manages the real-time
transmission of the control and patient samples to a host computer.
•
Use the Data Clean feature of the Online Settings window to prevent the transmission
of questionable sample results. The Transfer Results List dialog box of the
Review/Edit window indicates which samples failed this data check and were not
transmitted.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...