174
ADVIA 2400 Operator’s Guide
8 System Setup
Calibrating the touch screen monitor
Use this procedure to calibrate the touch screen at your workstation.
Calibrating the touch screen
1.
Right-select the
touch-screen
icon at the bottom right of the screen.
2.
Select
Align
.
3.
Select the targets that display at the screen.
4.
Select the check mark.
5.
The touch screen is calibrated.
Configuring serum indices
Use the following procedure to obtain the same flagging levels used by
Siemens when evaluating method serum interferences
•
Use the Qualit. set button in the Analytical Parameters (Serum) window to set the
five different flagging levels for each serum index.
•
Use the Item (Test) and Factor boxes in the Analytical Parameters (Serum) window
to designate the test for the determination of serum indices.
•
Use the Serum indices area on the System Specifications Set window to determine
when the serum blank assay is run.
Refer to the
System Configuration
section of the online
Operator’s Guide
for further
information.
Connecting to your host computer
1.
Have your laboratory computer professional connect the host computer as described
in the publication titled
LIS Interface Guide
.
2.
Use the Online Settings window to enter the communication parameters.
Use the Online settings window to determine which sample results and workorders to
transmit.
This feature allows you to transmit results when they are available. This is called
“real-time transmission.”
If you do not want to use real-time transmission, you can always transmit the results
as a batch using the Review/Edit window.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...