76
ADVIA 2400 Operator’s Guide
Loading patient samples
Loading patient samples on the STT
BIOHAZARD
LASER WARNING
Wear personal protective equipment.
Use universal precautions.
Do not stare into the beam.
NOTES
Before starting a run, you must load control and calibrator samples on the onboard
sampler.
To load patient samples while the system is sampling, select the SMP Pause button at the
Operation Panel or on the analyzer.
1.
Remove the sample tray evaporation cover.
2.
Remove any completed samples and dispose of them in accordance with laboratory
procedure.
3.
Make sure that on the Operation Panel, the Load Status Indicator line is green and
that SAMPLE LOAD OK displays
4.
Load samples while the sample tray is in the sampler, or remove the tray.
To remove the sample tray, release the safety latches at position numbers 49 and 70,
then grasp the sample tray by its handles and lift it out of the sampler well.
5.
Loading the samples.
a.
When using barcode labels:
•
Clean and properly position barcode labels on the sample tube.
•
Make labels face outward and visible through the slot in the sample-tube
holder.
•
Labels on tubes loaded in the inner ring (positions 43 through 84) must be
visible between the sample-tube holders in the outside ring.
b.
When not using barcode labels, you must load each sample into the sample
position number entered on the workorder.
c.
Load sample cups into a plastic adapter.
This adapter can hold two cup sizes. If the cup does not fit, remove the adapter
and try the other end.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...