138
ADVIA 2400 Operator’s Guide
6.
Fasten the connector.
7.
Using a pad soaked in alcohol, clean the outside surfaces of the filter holders and
hoses.
8.
Insert the filter hoses into the bottles, then fasten the caps.
9.
Return the analyzer to the READY mode.
10.
At the Operation Panel, select
Prime
.
11.
In the PRIME Set
dialog box, select
Prime 2
and enter
10
or more for the number of
times in all fields, and then select
Execute
.
Cleaning the pure-water bottle filter
Materials required:
• Ten 10R filters (REF 01160530,
PN 073-0035-01)
• One 18R filter (REF 01448895,
PN 073-0034-01)
• Hex wrench
Time: 15 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
Clogged filters create an insufficient flow rate and produce air bubbles.
NOTE:
A set of filters is included in the supplies kit. To avoid system down-time, replace
the filter with the one in the kit, resume operation, and then clean and store the removed
filter for the next scheduled maintenance.
1)
Pure-water bottle
Figure 5-24. Pure-water bottle
1.
Ensure that the instrument is in READY mode.
2.
Remove the silicon return hose (
1
) from the top front of the pure water bottle.
The filter is contained in a metal filter holder attached to the end of the return hose.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...