122
ADVIA 2400 Operator’s Guide
3.
Replace the reservoir cover.
4.
Replace the lamp access cover.
NOTE:
If adding coolant does not clear the lamp coolant warning, call your local
technical support provider or distributor.
Checking lamp energy
NOTES
Check the lamp energy after cleaning or replacing cuvettes, and after replacing the lamp.
When completing weekly maintenance, be sure to perform the procedures in the
following sequence:
1.
Weekly wash
2.
Lamp energy check (this procedure)
3.
Cuvette blank measurement
Materials required:
No materials required
Time: 15 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
IMPORTANT
Do not touch or turn the reaction tray at any time during the lamp energy check
procedure. The reaction tray should turn freely. If the reaction tray is shifted, repeat the
procedure, since a shift could result in an erroneous lamp energy reading.
1.
At the Menu Panel, select
Maint
, then select
Lamp Energy Monitor
.
The Lamp Energy Monitor window displays.
2.
Ensure the bottle at position #50 in Reagent Tray 1 (RTT1) contains deionized water.
3.
Select
Check Energy
.
The Lamp Energy Monitor dialog box displays.
4.
Type
50
in the RTT1 bottle posi. field, then select
3: 70 mL
for the Container
field.
5.
Select
Meas. Start
.
•
The reagent probe aspirates deionized water from RTT1 and dispenses it into
reaction cuvette #1.
•
The reaction disk rotates until cuvette #1 is in the detection position.
•
The Operation window displays Lumi.Check and then WAIT.
NOTE:
Perform steps 6–10 while in the WAIT state.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...