56
ADVIA 2400 Operator’s Guide
Methods 1–150 are predefined methods. The analytical conditions are predefined. Some
of the fields of the analytical conditions are grayed out and cannot be edited.
NOTE:
These are Siemens defined methods, only edit user-defined parameters.
Methods 151–200 are reserved for user-defined methods.
NOTE:
When you change and save your settings in the analytical parameter fields the
version number in the upper-right corner of the window moves up one numerical
increment and the date is updated to the current date, for example, from Ver.
2.0.0.yyyymmdd to Ver. 2.0.1.yyyymmdd (in this example the 2 is the software version
and the first 0 is the method parameter revision minor number).
System Test List window
Perform any of the following tasks at this window:
•
Add tests that are used regularly by your system. The tests on the list appear on other
windows
•
Assign a reagent tray position number for non-barcoded reagents or a reagent code
(R-Code) for barcoded reagents
CAUTION
To avoid incorrect reporting of results at the Review and Edit, the Daily QC
windows, and the Print Report form, you
must not modify the order
of the test
items on the System Test List window after initial setup. If you reposition any test
items on the System Test List window, results for samples already run could be
associated with the wrong test name.
Perform any of the following tasks at this window:
•
add tests to list
•
remove tests from list
•
set the test processing order
Analysis order oet window
Use this window to set the order in which tests on the list run. To run tests in the order
they displays at the System Test List window, you do not need to use this window.
•
print test list items
•
assign reagent codes or position numbers
Analytical Parameters (Serum) window*
Use this window to set the parameters for evaluating the serum indices (lipemia,
hemolysis, and icterus).
Perform any of the following tasks at this window:
•
define up to five different flagging levels of concern for each serum indices
•
select a test item to perform the serum indices analyses
•
enter the factor values required to calculate the serum indices result values
NOTE:
Operation of the serum indices feature is controlled in the Serum set area of the
System Specifications Set window.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...