192
ADVIA 2400 Operator’s Guide
Item Description
Assay
Measurement point
41 detection points/14 seconds in 10 minute reaction
Photometer
Concavity diffraction grating, rear spectroscopy system
Measurement wavelength
14 fixed wavelengths (340, 410, 451, 478, 505, 545, 571, 596, 658, 694,
751, 805, 845, and 884 nm), 1 or 2 wavelength calculation
Light source
12V, 50W halogen lamp, cooled by forced water circulation
Assay method
Colorimetric assays
•
Endpoint assays (EPA)
•
Reaction rate assays (RRA)
•
2-point rate assays (2PA)
•
3-item simultaneous measurement (parameter independent)
•
Prozone checking (antigen excess check)
•
Substrate depletion check with point forwarding option
•
Sample blank correction
Reassay mechanism
Automatic or manual (selectable)
Automatic correction
Blood serum blank, cell blank, measurement point change, sample volume
change in reassay
ISE
Method
Na, K, Cl ion selective electrolyte assay
Dilution measurement method
Analysis item
Simultaneous 3 item measurement of Na, K, and Cl
Electrode
Na: Crown ether membrane
K: Crown ether membrane
Cl: Super-layer solid molecule orientation membrane
ref: Sealed silver/silver chloride electrode
Throughput rate
Maximum of 600 tests/hour (200 samples/hour) for serum samples
Sample volume
22
µ
L
Dilution ratio
1:33 (approximate)
Reagent volume
Buffer: 2.7 mL/sample
Maintenance
Automatic maintenance
Automatic startup and automatic shutdown by timer for weekly
maintenance
Dimensions
Analyzer dimensions
1340(h) x 1711(w) x 934(d) mm
(52.8 x 68.44 x 37.36 in.)
Weight
630 kg (1389 lb.)
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...