140
ADVIA 2400 Operator’s Guide
CAUTION
To assist with removal as well as to avoid crimping these thin filter hoses, remove
only 3 or 4 hoses from the water bottle at a time.
6.
First unfasten the filter holder from the end of each tube, and then remove the filter.
7.
To clean the 10R filter:
a.
Place the filter in a beaker filled with a freshly made 10% solution of household
bleach and water.
b.
After 30 minutes, remove the filter and rinse it with deionized water and replace
onto the filters.
CAUTION
Ensure that the filters are properly positioned with the filter holder to avoid filter
shift.
8.
Using a lint-free towel soaked in 5% bleach solution, clean the outside surfaces of the
filter holder and hose.
9.
Insert the silicon filter hose into the water bottle.
10.
Insert the ten Teflon filter hoses in the tank, 3 or 4 at a time.
11.
Prime the lines:
a.
At the Operation Panel, select the
Prime
button.
b.
In the PRIME Set dialog box, select
PRIME 2
and type
10
or more for the
number of times in all fields.
c.
Select
Execute
.
Replacing the reaction and dilution cuvettes
Materials required:
• 13 sets of reaction cuvettes (sample
cell RRV, single cuvette set,
REF 05024992, PN 073-0023-02)
• 6 sets of dilution cuvettes (sample cell
DTT, single cuvette set,
REF 05049669, PN 073-0022-01)
Time: 20 minutes
Analyzer mode: OFF
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
Replace the 20 sets of reaction (RRV) cuvettes and 6 sets of dilution (DTT) cuvettes once
every four months.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...