108
ADVIA 2400 Operator’s Guide
5.
Return the analyzer to the READY mode.
6.
Verify that the analyzer mode is READY before performing any further actions.
NOTE:
If an overflow error message appears, there is probably water on the sensor.
Dry the sensor.
To clear the alarm message, at the Operation Panel, select the audible alarm icon
.
Checking reagents and system solutions
Refer to page 69,
Checking the availability of the reagents and wash solutions
, in the
Operating the System section in this manual or, for more detail, refer to the online
Operator’s Guide.
Inspecting and cleaning the reaction (WUD) and dilution (DWUD)
cuvette washers
Materials required:
• Lint-free towel
• Alcohol prep pads or lint-free towels
and 5% bleach solution
• 4-mm hex wrench
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1.
Inspect the exterior of the reaction cuvette washer (WUD) and dilution cuvette
washer (DWUD) tubing for cleanliness.
2.
Check the WUD and DWUD for leaks.
TIP
Perform this inspection in addition to the startup, shutdown, and weekly washes, to
keep the WUD and DWUD from clogging.
In the event of a clog, call your local technical support provider or distributor for
assistance.
Figure 5-4. Cuvette wash stations
1)
Dilution Cuvette Wash
Station (DWUD)
2)
Reaction Cuvette Wash
Station (WUD)
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...