Appendix
191
Item Description
Minimum sampling volume
2
µ
L (0.1
µ
L increments)
Special dilution
Diluted sample can be rediluted directly from tray.
Dilution
Dilution tray (DTT)
Turntable system
Number of cuvettes
120 (6 sets of 20 cuvettes)
Maximum sample volume
300
µ
L
Dilution cuvette dead volume
35
µ
L
Dilution ratio
From 1 to 1:5625 (1:75 x 1:75)
Reaction cuvette material
Plastic
Reagent
Dispensing system
2-reagent capability, 2-probe system
Trays
Two trays, each holding 50 containers
Multiple reagent pack loads – 5 reagents per method with automatic
rollover upon depletion
Reagents on each tray are cooled to between
6
°
C and 14
°
C
Container capacity
20 or 70 mL. (40 mL with adapter)
Refrigerator Recirculated
liquid
Reagent volume/item
10 to 100
µ
L (0.1
µ
L increments)
Dilution reagent volume
10 to 100
µ
L (0.1
µ
L increments)
Reagent inventory
Computed by the liquid-level sensing
Barcode
Interleaved 2 of 5. JSCLA standard.
Reaction
Reaction tray (RRV)
Turntable system
Number of cuvettes
340 (20 sets of 17 cuvettes)
RRV cuvette material
Plastic
Reaction liquid volume
60 to 180 mL
Stirring system
Rotation and reciprocation
Strong and weak stirring options are available.
Stirring immediately after additions of R1 and sample and R2
Mixer 1 is used for R1; Mixer 2 is used for R2.
Reaction times
3, 4, 5, 10 minutes
Extended reaction times: 15 & 21 minutes
Reaction temperature
37
°
C
Temperature regulation:
±
0.1
°
C
Reaction tank
Inert liquid circulation system
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...