150
ADVIA 2400 Operator’s Guide
Priming the system
1.
At the Menu Panel, select
Maint.
, then select
User maint
. (For additional details,
refer to Using the User Maintenance window.).
2.
At the User Maintenance window, in the Position Probes for Routine Cleaning area,
select
Start
, then select
Yes
when prompted.
3.
Ensure that the probe is perpendicular to the arm and centered over the cuvette.
4.
If not centered, call your local technical support provider or distributor.
5.
At the Operation Panel, select
Initialize
to return the probes back to home (over the
wash cups).
6.
At the Operation Panel, select
Prime
,
PRIME 2
, and then
Execute
to ensure proper
water flow through the probe.
NOTE:
Make sure that no water is leaking from the joint connector (
4
).
Replenishing the RRV (reaction) bath oil bottle
Materials required:
• RRV (reaction) bath oil
(REF 09323099, PN B01-4180-01)
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
1)
ISE
buffer bottle
2)
RRV (Reaction) bath oil bottle
3)
Isotonic saline diluent bottle
4)
Cuvette detergent bottle
5)
Cell conditioner bottle
Figure 5-33. RRV bath oil bottle
CAUTION
Do
not attempt to clean the RRV bath oil bottle (
1
) with water; RRV bath oil and
water do not mix.
1.
Unscrew the filter cap (
3
) at the front of the RRV bath oil bottle (
1
), then pull up the
tube with the filter.
2.
Disconnect the RRV bath oil bottle level sensor connector (
2
), then turn the
connector counter-clockwise and pull it out.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...