Maintenance
103
Daily maintenance
Inspecting and cleaning the probes
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions
.
Materials required:
• Phillips screwdriver
• Alcohol prep pad or lint-free towels and
5% bleach solution
Time: 10 minutes
Analyzer mode: READY
1.
Visually inspect each probe daily.
1)
Dilution Probe (DPP)
2)
Sample Probe (SPP)
3)
Reagent Probe 1 (RPP1)
4)
Reagent Probe 2 (RPP2)
Figure 5-1 Location of probes
•
Replace any clogged probes. See
Replacing the Probes
in the online Operator’s
Guide.
•
Clean all probes daily (proceed to step 2).
TIP
Perform the Shutdown Wash and Weekly Wash as scheduled, to prevent the probes
from clogging.
2.
Clean each probe using one of the 3 methods described below:
Cleaning the probes using automatic advance probe motion
1.
At the Menu Panel, select
Maint
, then select
Manual Operation
.
2.
At the Manual Operation window, double-select the
code
for the probe you want to
move as follows:
Probe Code
dilution probe
3.DPPLR
sample probe
16.SPPLR
Reagent probe 1
37.RPPLR-1
Reagent probe 2
49.RPPLR-2
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...