System Overview
19
PC (back view)
1)
PC power connector
2)
Mouse connection
3)
Keyboard connector
4)
Communication port 1 (LIS-TDC)
5)
Printer connector
6)
LCD monitor connection
7)
Analyzer Ethernet connection
8)
Communication port 2 (URH)
9)
Sleep ITF board potentiometer
Figure 1-10. PC back view
Sampling and analysis
Sampling mechanisms
The sample probe (SPP) aspirates sample from the dilution tray (DTT) and dispenses it
into reaction tray (RRV) cuvettes for analysis, according to specified conditions. The
sampling pump (SP) handles the aspiration and dispensing functions.
Sampling operation
At initialization, the SPP moves (in the up position) to the RRV cuvette, then stops above
the wash port. The pumps stop when the SPP is positioned over the RRV cuvettes.
Each sampling cycle consists of these steps:
1.
The SPP moves to the aspiration position of the DTT and aspirates the sample.
2.
The SPP moves back to the wash port, where deionized water washes the SPP’s
outside surface.
3.
The SPP moves to the RRV and dispenses the sample into a cuvette.
4.
The SPP returns to the wash port, where degassed water washes the inside and
outside of the SPP.
The SPP is now ready for another cycle.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...