118
ADVIA 2400 Operator’s Guide
6.
At the WASH Set window, define the WASH2 container positions as follows:
a.
Select
WASH2
.
b.
Select
2
for Cycles.
c.
Enter
49
in the CTT cup position 1st time field and
50
in the CTT cup position
2nd time field.
d.
Enter
49
in the RTT1 and RTT2 cup positions 1st time fields and
50
in the RTT1
and RTT2 cup positions 2nd time fields.
e.
Select
Execute
.
Performing additional ISE electrode washes
Materials required:
• ISE Detergent Solution
(REF 01307361, PN B01-4174-01)
Time: 5 minutes
Analyzer mode: Manual Operation
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
CAUTION
Use only Siemens-recommended
ancillary reagents to avoid
erroneous test results.
WARNINGS
•
Read and follow the cautions on
the box and label before
handling any reagents. Some
method and system reagents are
classified as hazardous material.
•
Avoid contact of method or
system reagents with skin and
eyes. These materials can cause
infection and burns. Wear
suitable protective clothing,
gloves, and eye/face protection.
In case of contact with eyes,
rinse immediately with plenty
of water and seek medical
advice. In case of contact with
skin, immediately wash with
soap and water.
•
Avoid inhaling chemical
vapors. If you inhale chemical
vapors, promptly leave the area
and seek fresh air.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...