110
ADVIA 2400 Operator’s Guide
Figure 5-6. Wiping the wash stations with lint-free towels
6.
Reinstall the wash head:
a.
Replace the wash head using the alignment pins located on either side of the
retaining screw, then tighten the 4-mm hex screw.
b.
Ensure that all tubes are securely connected.
c.
Remove the toweling.
d.
Ensure that each nozzle is centered above the corresponding cuvette.
7.
At the Operation Panel, select Initialize and verify the the DWUD and WUD are in
the up position and that the analyzer is in READY mode.
8.
Verify the wash head nozzles are correctly centered in the cuvettes:
a.
At the Menu Panel, select
Maint
, then select
Manual Operation
.
For additional information concerning the Manual Operation window, refer to
the Manual Operation window in the online Operator’s Guide.
b.
At he Manual Operation window, double-select
14.DWUD
or
23.WUD
.
c.
Select
Move
to slightly lower the washer nozzles, then verify that they are
correctly positioned.
If not, call your local technical support provider or distributor.
d.
Verify the nozzles are correctly centered.
e.
At the DWUD or WUD window, select
Init
., then select
Exit
to raise the washer
nozzles.
9.
At the Operation Panel, select
Initialize
, then
verify the DWUD and the WUD are in
the up position and the instrument is in the READY
state.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...