164
ADVIA 2400 Operator’s Guide
5.
Select
Re-Start
, then select
OK
.
6.
If possible, repeat the task that you were performing prior to the power failure and
verify that the data was stored.
7.
If reagent was dispensed, you must perform a Weekly WASH2 before resuming
operation.
Recovering from an unexpected power outage (after power returns,
when power was unexpectedly lost while system was on)
1.
If the Startup window is open, select
Shutdown
and perform the normal shutdown
operation.
2.
Turn off the workstation power and turn the
Operate/Standby
switch on the analyzer
to
Standby
.
3.
Wait approximately 20 seconds.
4.
Perform the normal startup operation and open the Startup window.
5.
At the Startup window, enter a password.
6.
Select
Re-Start
, then select
OK
.
7.
If possible, repeat the task prior to the power failure and verify that the data was
stored.
8.
If reagent was dispensed, you must perform a Weekly WASH2 before resuming
operation.