80
ADVIA 2400 Operator’s Guide
Running calibrators and reagent baselines (RBL)
1.
Select
Multipnt.smp. Analyze
, then select
TT No. 98
or
99
to select the STT tray
number.
2.
Select
One-pnt.smp. Analyze
.
You can select both calibration types.
3.
Select
Ordinary calib
. or
Special calib
.
4.
To remove any tests you do not want calibrated in this run, select
Temp.item select
.
5.
To remove any calibrators you do not want used in this run, select
Temp.sample
select
.
To run only a reagent baseline (RBL) or blank rate, select only the blank solution.
Running control samples
1.
Select
Control smp. Analyze
.
2.
To remove any tests you do not want run on the controls, select
Temp.item select
.
3.
To remove any controls you do not want used in this run, select
Temp.sample
select
.
From STT
1.
Make sure the samples are loaded.
2.
In the Ordinary sample area, next to Routine smp, select
Analyze
.
This activates other fields.
3.
To specify how the samples are identified, in the Ordinary sample area, next to
Analyze mode, select
Barcode
or
Cup posi.
If you select
Barcode
, the system scans the entire STT tray for barcodes.
You can choose to scan for a single tray position or for a range of positions as
follows:
4.
In the boxes to the right of Analyze, enter the range of positions you want to scan.
For example: If you enter 50 and 60. The system scans for barcodes within this range
and will aspirate only from positions at which it found barcodes.
If you want the system to aspirate sample from a specific position or specific range of
positions, select
Cup posi
.
Enter the tray number.
In the boxes to the right of Analyze, enter the position or range of positions from
where you want the system to aspirate samples.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...