126
ADVIA 2400 Operator’s Guide
Monthly maintenance
Cleaning the turntable interiors (STT/CTT and RTT)
Materials required:
• Phillips screwdriver
• Lint-free towels
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
NOTE:
Use the 2 procedures that follow to clean inside the STT/CTT housing and the
RTT1 and RTT2 refrigerated housing to remove accumulated sample, reagent, dust, and
other materials.
Cleaning the inside of the STT/CTT housing
1.
Remove the Calibrator/Control Tray loader (CTT):
a.
Lift the standard cover from the loaders.
b.
Pull up on the two Nylatch fasteners (
3
) securing the CTT Tray loader in place.
c.
Lift out tray loader by the center handle (
4
).
2.
Remove the Sample Tray loader (STT):
a.
Lift the Sample Tray evaporation cover.
b.
Pull up the two fasteners (
5
) securing the STT Tray loader in place.
c.
Lift out the STT tray by the two metal handles.
3.
Using lint-free towels, wipe the interior of the STT and CTT housings.
1)
CTT Tray
2)
STT Tray
3)
CTT Nylatch Fasteners - 2 places
4)
CTT Handle
5)
STT Nylatch Fasteners - 2 places
6)
STT Handles - 2 places
7)
Locator Screw
Figure 5-16. Components of CTT and STT trays
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...