ADVIA 2400 Operator’s Guide
This information is used by the Reagent Inventory window.
Perform any of the following tasks at this window:
•
set the reagent container type, lot number, and expiration date
•
start the Days Remaining counter for non-barcoded reagent containers
•
print the settings
•
clear the settings
Active Test List window*
Use the Active Test List window to temporarily deactivate a method, so that method-
related alarm messages (for example, missing reagents) are not generated. You would use
this function if, for example, for tests that you do not run every day.
After you make your selections, activating or deactivating a method, you must initialize
the system for the changes to take effect.
The Active Test List window displays all tests loaded on the system, but the deactivated
or unchecked tests are grayed out.
NOTE:
At the Active Test List window, you must select all the tests you expect to run. If
a test is not selected, the system skips that test. An error message does
not
display in the
Alarm message box on the Operation Panel, but is recorded in the error message log.
Active is checked if the test is active.
Test Name displays the name and number of the test.
R1 Pos No displays the position of R1 on the reagent tray.
R2 Pos No displays the position of R2 on the reagent tray.
Reagent Information window*
Use this window to review and edit reagent information, such as on-system stability,
calibration frequency, and fill volumes for barcoded reagents and to enter information for
non-barcoded reagents.
The Reagent Inventory and Reagent Container Set windows use this information.
Assay Name
is the test method name.
R-Code
is the reagent code for each Siemens barcoded reagent container or a 5 digit user
defined code for non-barcoded reagents.
Days Remaining
specifies the on-system reagent stability.
Comment
is a user entered remark or reminder.
R1
and
R2
Fill
Volume
are the fill volume in mL for reagent containers R1 and R2.
NOTE:
For user-defined methods, the Assay Name, R-code, Days Remaining and R1-
R2 Fill Volume must be entered correctly. R2e is not used.
After system installation, all default information for Siemens-defined methods appears in
this window.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...