Operating the System
69
Operating condition
Normal
indicator
Abnormal indicator and corrective
action
Diluent vol.
Indicates the status of the liquid volume of
diluent (saline solution). Insufficient liquid
is detected by a liquid level sensor in the
bottle.
OK
NG
Fill the diluent bottle with 0.9% saline
solution.
Make sure the sensor is in the proper
position.
Cuv. Wash Vol.
Status of the liquid volume of cuvette wash
solution. Insufficient liquid is detected by a
liquid level sensor in the bottle.
OK
NG
Fill the cuvette wash solution bottle.
Make sure the sensor is in the proper
position.
Cuv. Cond. Vol.
Status of the liquid volume of cuvette
conditioner. Insufficient liquid is detected
by a liquid level sensor in the bottle.
OK
NG
Fill the cuvette conditioner bottle. Make
sure the sensor is in the proper position.
Conc.wast.tank
(if installed)
Status of the volume of the optional
concentrated waste tank. A sensor monitors
whether the tank is filled with waste.
OK
NG
Empty the concentrated waste tank.
If there is still a problem, check the tubes
for obstructions.
Waste tank
Status of the volume of the optional waste
tank. A sensor monitors whether the tank is
filled with waste.
OK
NG
Verify that the waste tubes are free of
obstructions. If the problem persists,
contact your local support provider or
distributor.
3.
Checking the availability of the reagents and wash solutions.
Visually check the system reagents.
a.
Perform a prime after replacing any system ancillary reagents.
b.
At the Operation Panel, select
Prime
.
c.
Select
PRIME 2
, then select
Execute
.
d.
Visually check the ISE reagents.
e.
Perform a prime after replacing the ISE Buffer:
(1) At the Menu Panel, select
Maint
.
, then select
ISE Operation
.
(2) In the Times box of the Bufferprime area, type the number of prime cycles
you want (
20
is recommended).
(3) Select
Execute
.
f.
Visually check the controls and calibrators on the calibrator/control tray.
g.
Visually check the wash solutions.
h.
Visually check the lamp coolant level.
Summary of Contents for ADVIA 2400
Page 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Page 12: ...12 ADVIA 2400 Operator s Guide ...
Page 64: ...64 ADVIA 2400 Operator s Guide ...
Page 98: ...98 ADVIA 2400 Operator s Guide ...
Page 165: ...Maintenance 165 ...
Page 167: ...Troubleshooting 167 ...
Page 173: ...System Setup 173 ...
Page 177: ...System Setup 177 ...
Page 181: ...Appendix 181 ...
Page 182: ...182 ADVIA 2400 Operator s Guide ...