- 92 -
Upon selecting a series, press one of the following buttons determining how the
differential weighing is to be carried out:
Weighing A
Press to start weighing initial mass.
Tare (T)
Press to start sample container taring.
Weighing (T+A)
Press to start sample container taring which is next followed automatically by
sample weighing (non-separable operations).
Weighing B
Press to start weighing final mass. The button is enabled only for a series
containing samples for which weighing A has been carried out. If there are no
such samples, the operation is disabled.
Weighing A
Upon pressing Weighing A button, the balance
searches samples list in order to find the first
sample for which weighing <A> is still to be
carried out. If there are no such samples the
balance displays a message informing you
that the operation is disabled.
If the operation is enabled then new data on
initiated process is displayed in the grey
workspace.
You are guided through the process by means
of prompts visible either in the workspace or
bottom bar.
Load the weighing pan with a sample and
press <
> button.
<Ambient condition> parameter set to <Value>
parameter a window for entering temperature
and humidity values is displayed. For
<Ambient condition> parameter set to
<ONLINE> parameter a window for
confirmation of values taken from ambient
conditions module is displayed.
CAUTION:
If <Ambient conditions> parameter for the
differential weighing mode is set to <No>
value the software skips this part of the
process.