- 130 -
3. Date [process performance date]
4. Tare [determined packaging tare value]
5. S
[average standard deviation]
6. 0.25 T1 [the value of a criterion for process result]
7. Measurements quantity [quantity of carried out tare value measurements]
8. Operator [name of an operator carrying out the process]
9. Method [control method
(regulated by the act or by internal criteria)]
30.10. Pipettes
Pipettes database contains data of pipettes that are to be calibrated.
Procedure:
•
Enter <
Databases> submenu and press <
Pipettes > entry.
•
Press <
Add> button to add a new pipette, or
•
Select already existing pipette.
List of parameters defined for pipettes:
1. Name
2. Code
3. Model
4. Tip
5. Volume type
6. Nominal volume
7. Minimum volume
8. Channels quantity
9. Type
10. Tested volumes
30.11. Pipette Calibration Reports
Pipette reports provide information about carried out pipettes calibrations. In case of
multichannel pipettes, reports are generated for each channel separately. Each report can
be previewed, searched by date or exported and printed out.
Procedure:
•
Enter <
Databases> submenu and press <
Pipette calibration reports> entry.
•
Select report, if not visible, scroll the list using navigation buttons.
•
Report is named by date and time; e.g. 2012.03.12 11:12:15.
CAUTION:
It is possible to use ‘search report’ option.
List of pipette calibration reports data
:
1.
Pipette
[pipette name]
2.
Serial no.
[serial no. entered during calibration]
3.
Channel no.