- 173 -
x
- balance series (inserted in between inverted commas), with general balance type in front.
Example:
Command:
BN <CR><LF>
- give module type
Response:
BN_A_”AS”
Balance type XA 4Y
Give max capacity
Format:
FS <CR><LF>
Response options:
FS_A_”x” <CR><LF>
- command understood, response: module max capacity
FS_I <CR><LF>
- command understood but not accessible at this moment
x
- Max value of reading units (in between inverted commas)
Example:
Command:
FS <CR><LF>
- return Max capacity
Response:
FS_A_”220.0000”
- Max capacity: ”220 g”
Give program version
Format:
RV <CR><LF>
Response options:
RV_A_”x” <CR><LF>
- command understood, response: program version
RV_I <CR><LF>
- command understood but not accessible at this moment
x
- program version (in between inverted commas)
Example:
Command:
RV <CR><LF>
- return program version
Response:
RV_A_” 1.1.1”
- program version: ”1.1.1”
Set AUTOZERO function
Format:
A_n <CR><LF>
Response options:
A_OK <CR><LF>
- command carried out
A_E <CR><LF>
- error in-course of command execution, no parameter or incorrect
format
A_I <CR><LF>
- command understood but not accessible at this moment
n
- parameter, decimal value determining autozero settings
n
→
0 – autozero off
1 – autozero on
CAUTION:
Command changes settings for a current working mode.
Example:
Command:
A_1<CR><LF>
- turn autozero function on
Response:
A_OK<CR><LF>
- autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
Set ambient conditions state
Format:
EV_n <CR><LF>
Response options:
EV_OK<CR><LF>
- command carried out
EV_E <CR><LF>
- error in-course of command execution, no parameter or incorrect