Procedure:
Press entry with test name, a message box is displayed.
You can:
•
Start subsequent AUTOTEST GLP process,
•
Preview results of completed autotests, and export all saved test results as file with
*.csv extension,
•
Delete all saved test results.
You can abort the process at any moment by pressing <X> button.
Results of carried out autotests are displayed in a table (each line features date and time of
autotest procedure and its result).
To preview particular autotest data, press respective table line.
In order to print a result of a single autotest, enter its details (see above) and generate a
printout by pressing <Print> key located on the top of the display.
Autotest results can be exported by pressing <Export> entry in the window with all saved
reports. Data is sent to a USB flash drive connected to balance USB port as a file with *.csv
extension.
34.12. Start-Up Logo
(option enabled only for authorized mass comparator operators)
The parameter enables you to change logo that is displayed on balance start-up.
34.13. Export of System Events
(option enabled only for authorized mass comparator operators)
The parameter enables you to generate a special type of a file that is automatically saved to
a USB flash drive, plugged to USB port. The file helps RADWAG service to diagnose cause
of problems occurring during balance operation.
Procedure:
•
Plug a USB flash drive to mass comparator’s USB port.
•
Enter <
Misc.> submenu.
•
Select <Export system events> parameter.
•
The software generates a file and automatically saves it on the USB flash drive.
•
The generated file has to be sent to RADWAG.
35. UPDATE
Update function enables update of:
•
Help files available from the operator level.
•
Application.
•
Metrological parameters: main board (administrator exclusively).
Update takes place automatically by loading data from a USB flash drive connected to
balance USB port.
Procedure:
•
Prepare a USB flash drive with update file. Required file extension:
*.lab2.
•
Connect the USB flash drive to balance USB port.
•
Select <Application> entry.