- 28 -
10.8. Report Printout
<
Report printout> parameter determines whether or not an adjustment report is to be
automatically printed upon adjustment process completion.
In order to set automatic report printout, go to <
Report printout> parameter and select
<YES> value.
10.9. GLP Project
GLP is one of many methods for documenting work in accordance with adopted quality
system. Data selected for printing is printed on each release of an adjustment report.
Information and signs to be used in a GLP report:
adjustment (adjustment type)
working mode (working mode name)
date, time
operator
balance type
balance S/N
level status
nominal mass
current mass
difference
temperature
empty line
dashes
signature
non-standard printout
10.10. Adjustment History
Adjustment history contains data on all carried out adjustment processes. The record is
carried out automatically. Each record on adjustment comprises basic data on completed
process. This menu enables displaying the list of completed adjustment processes. Each
report is printable.
In order to print an adjustment report, enter <Adjustment> submenu, next enter <Adjustment
history>, select adjustment to be printed. Upon displaying record details, press <
key located on the top of the display.
CAUTION:
If the balance memory is full, the oldest record gets automatically deleted.
If internal procedures of an organization require maintaining complete documentation
concerning all carried out adjustment processes, then the list with adjustment records has to
be printed and archived periodically.
Searching for Adjustment Record
The module enables searching for a specific completed adjustment record. To do that, press
search button and enter date of an adjustment process that is to be found.