Press <
> button to confirm ambient
conditions values, differential weighing home
screen is displayed, you are prompted to
unload the weighing pan.
Unload the weighing pan and press <
> button.
The balance searches samples list again in order to find the first sample for which
weighing <A> is still to be carried out. Follow the above procedure and carry out
weighing of the remaining samples.
In order to abort the process press <
> button.
Weighing T
Upon pressing Weighing T button, the balance searches samples list in order to find
the first sample for which weighing <T> is still to be carried out. If there are no such
samples the balance displays a message informing you that the operation is disabled.
If the operation is enabled then new data on initiated process is displayed in the grey
workspace.
You are guided through the process by means of prompts visible either in the
workspace or bottom bar.
Load the weighing pan with a sample container and press <
> button.
Upon pressing <
> button you are prompted to unload the weighing pan.
Unload the weighing pan and press <
> button. The balance searches samples list
again in order to find the next sample for which weighing <T> is still to be carried out.
Follow the above procedure and weigh the remaining containers.
In order to abort the process press <
> button.
Weighing T+A
Upon pressing Weighing T button, the balance searches samples list in order to find
the first sample for which weighing <T> is still to be carried out. If there are no such
samples the balance displays a message informing you that the operation is disabled.
If the operation is enabled then new data on initiated process is displayed in the grey
workspace.
You are guided through the process by means of prompts visible either in the
workspace or bottom bar.
Load the weighing pan with a sample container and press <
> button.
Container mass value is assigned to a sample as tare value, balance indication is
zeroed, and you are prompted to place the sample in the container. Follow the prompt
and press <
> button.
For <Ambient condition> parameter set to <VALUE> parameter a window for entering