- 65 -
CHECKWEIGHING
Checkweighing function facilitates control of sample mass, mind that Min and Max
thresholds, given in [%], are required.
Procedure:
1.
Press
button (Checkweighing thresholds).
2.
Press Min threshold button and enter its value [%].
3.
Press
button to confirm.
4.
Press Max threshold button and enter its value [%].
5.
Press
button to confirm.
Caution: Max threshold value has to be greater than Min threshold value.
DOSING
Dosing function requires target value to be specified (it must be given in [%]) along with
dosing tolerance, i.e. permi/- surplus or shortfall.
Procedure:
1.
Press <
button (Target value).
2.
Enter the target value in [%].
3.
Set tolerance value, if required.
4.
Press
button to confirm.
5.
Press Min threshold button and enter its value [%].
6.
Press
button to confirm.
19.3. Bar Graph
The DOSING and CHECKWEIGHING functions are aided by a graphic indication, i.e. bar
graph. Example of simultaneous operation of DOSING and CHECKWEIGHING functions is
presented below.
•
<
>Checkweighing thresholds settings: Min= 90%, Max = 110%.
•
target value = 105%; tolerance = 5% <
>,
•
reference mass = 19.986 g <
>.