- 96 -
If tare is assigned to a sample, then its value is displayed with ‘-‘ sign. Load the
weighing pan with a sample (if tare is assigned load the sample in a container) and
press <
>.
For <Ambient conditions> parameter set to <Value> option, a window for entering
temperature and humidity values is displayed. For <Ambient condition> parameter set
to <ONLINE> option a window for confirmation of values taken from ambient
conditions module is displayed.
CAUTION:
If <Ambient conditions> parameter for the differential weighing mode is set to <No>
value the software skips this part of the process.
Press <
> button to confirm ambient conditions values, differential weighing home
screen is displayed, you are prompted to unload the weighing pan.
Unload the weighing pan and press <
> button. The balance searches samples list
again in order to find the first sample for which weighing <B> is still to be carried out.
Follow the above procedure and carry out weighing of the remaining samples.
In order to abort the process press <
> button.
Upon completing measurements of final mass for the samples in a series, you can check the
results in the series database. To check the results, enter the database, select a series, then
select samples and a specific sample for which <Weighing B> has been carried out.
Particular weighing preview mode provides you with values. Numbers referring to weighing
<B> inform on the measuring cycle.
Data on a specific series can be printed or exported to a file.
25.5. Copy Tare
Copy Tare option enables copying once selected tare value (in a series) to all samples which
do not have tare value assigned, and for which differential weighing process has not been
completed (status other than Weighing B).
Procedure:
Press