Autozero: Dosing
Last digit
Units
Enter this menu to declare the start unit, the supplementary unit, 2
custom units and to enter the value of gravitational acceleration force in
module’s operation place.
13.2.1. Settings
Name
Upon entering this parameter, a message box with an on-screen
keyboard is displayed. Enter the name of a profile and press
button to confirm. The name is assigned to the profile.
Default working mode
Upon entering this parameter you can select a specific working mode
that is to be set as default. For <None> value set, upon selecting the
profile, the balance activates the most recently used working mode.
13.2.2. Working Modes
Upon entering Working Modes parameter, a window containing all accessible working
modes is displayed. Parameters are set for each working mode independently. Particular
profile when activated runs with given set of parameters.
You can enter the following data for each working mode:
Settings:
contains specific parameters relating to a working mode and universal settings, such
as: result control, tare mode, automatic footer printout, printout mode, printout.
Quick access buttons:
declaring quick access buttons to be displayed.
Information:
declaring information to be displayed.
Printouts:
declaring type of a printout or defining a non-standard printout.
13.2.3. Readout
The function enables you to adjust balance operation to ambient conditions (filter settings) or
your individual needs. <Readout> menu contains the following elements:
FILTER (function unavailable for the following profiles: Fast, Fast
dosing, Precision)
Each measurement signal before being displayed is electronically
processed in order to provide correct parameters specific for stable
measurement result, i.e. ready to read.
You can determine signal processing rate, there are five different filters for you to
select from. Range:
•
very fast, fast, average, slow, very slow.
While setting the filter consider the actual operating conditions. In case of very good
operating conditions select average or fast filter; in case of rough conditions select
slow or very slow filter.
CAUTION:
for precision balances set: very fast
÷
or average filter,
for analytical balances and microbalances set:
÷
very slow filter.