- 55 -
Vibrations are recorded in the database, this allows to analyse history of ambient conditions
quality.
Effective value of vibrations registered by the balance over a declared time interval is saved
(from the moment of the last record) – Setup/Ambient conditions/Ambient conditions
recording interval.
By default the interval is set to 10 min, this value may range between 1 min - 60 min. Record
analysis allows the user to select optimal time of the day for the measurements performance,
i.e. time when the environmental conditions influence the vibrations the least. Mind that there
are places where the most convenient working conditions occur after-hours.
Along with each measurement, data on ambient conditions alerts and level of vibrations, that
occurred in the course of the measurement, is saved.