- 156 -
In case of verified instruments, the settings are hidden and the balance operates in
accordance with factory settings, i.e. <Active with lock>. Weighing process is possible only if
balance is properly levelled.
Procedure:
•
Enter <
Misc.> submenu.
•
Select <
Level control> parameter, an edit box is displayed.
•
Select one of available options:
a. None – level indicator not displayed, the balance does not control level status,
b. Active – level indicator displayed, level status is signalled by colour (green
→
OK, red
→
levelling required),
c. Active with lock – level indicator displayed, level status is signalled by colour
(green
→
OK, red
→
levelling required). For red level indicator ‘no Level’ alert is
displayed - weighing function is blocked.
CAUTION:
For detailed information on mass comparator levelling refer to section 10.3 of this user
manual.
34.9. Proximity Sensors Sensitivity
The parameter is adjustable between 0 – 9, and it determines the distance from which the IR
proximity sensors react.
Default value ranges between 5 and 7.
Procedure:
•
Enter <
Misc.> submenu.
•
Select <Proximity sensors sensitivity> parameter, an edit box is
displayed.
•
Select one of available options.
34.10. Proximity Sensors Timeout
Parameter for setting timeout of proximity sensors expressed in [ms].
Default value: <0>.
When you use the option of motion recognition set <Proximity sensors timeout> parameter to
<
500ms
> value.
Procedure:
•
Enter <
Misc.> submenu.
•
Select <Proximity sensors timeout> parameter, an on-screen keyboard is displayed.
•
Enter <500> value.
•
Confirm. Menu is displayed.
34.11. Autotest
< <AUTOTEST> function has been designed to help you assess balance operation
and to diagnose the cause of weighing errors exceeding the maximum permissible
values.
AUTOTEST facilitates regular optimisation of balance settings which is necessary in