QIAsymphony SP/AS Consolidated Operating Guide 12/2017
119
Loading reagent slots
Important
Ensure that the correct labware is used. Use of labware that is different to that
defined in the
Loading Information
screen may result in an error during
preparation or transfer of the master mix. This could result in damage to the
QIAsymphony AS.
To load a reagent adapter with reagents, follow the steps below.
1.
Open the “Eluate and Reagents” drawer.
2.
In the
Assay Setup/Loading information
screen, press the first “Reagents” slot (shown in
yellow). The detailed loading information for the slot is shown.
3.
Place the appropriate precooled reagent adapter onto the defined “Reagent” slot.
4.
Press the “Reagent” slot(s) to view detailed information about the required reagents, tubes,
and corresponding volumes. The
Loading Reagents
screen appears. A schematic of the
reagent adapter that will be used is displayed on the screen.