QIAsymphony SP/AS Consolidated Operating Guide 12/2017
145
Loading a normalization rack (only for assay run with normalization)
To load a normalization rack, proceed as follows:
1.
If not already open, open the “Assays” drawer. Temporary cooling for defined slots is started.
2.
In the
Assay Setup/Loading information
screen press the
Normalization
slot (highlighted
yellow).
Detailed loading information for the slot is shown.
3.
Place the empty normalization rack in the appropriate
adapter on slot 6
4.
Press
Load
. The
Assay Setup/Loading information
screen
reappears
The loaded slot is now highlighted blue.
5.
Leave the “Assays” drawer open to load disposable
filter-tips (see “Loading disposable filter-tips” on
page 122).
Important
Ensure that the appropriate adapter is used with the normalization rack.
Important
Do not load partially filled normalization racks.
12.4.9
Checking cooling temperatures
For instructions about how to check cooling temperatures, see Section 12.3.3.
12.4.10
Starting an independent run
Wait until the cooling positions have reached their target temperatures (i.e., when they appear
green in the assay setup
Overview
screen).