QIAsymphony SP/AS Consolidated Operating Guide 12/2017
163
Error code
Description
Comments and suggestions
75111
The adapter for the selected rack
type
QIA#19588 *EMTR
is not
configured. If available, configure the
adapter using the
Configuration
menu. If the adapter is not available
select a different rack type.
The selected rack type requires an adapter that has not been
configured in the list of available adapters and holders.
If you have the adapter type that is required, update the list of
available adapters and holders. See Section 6.1.3 of the
QIAsymphony SP/AS User Manual — General Description
for
more information.
If you do not have the required adapter type, change the rack
type assignment according to the types of adapters that are
available.
75403
The following Assay Parameter Sets
in the work list are unknown: X.
The work list contains Assay Parameter Set(s) that are not
available on the QIAsymphony SP/AS instruments.
If the Assay Parameter Set(s) are not required for the assay
run, continue and select the required assay(s).
If the Assay Parameter Set(s) are required for the assay run,
cancel the assay definition process. Transfer the required
Assay Parameter Set(s) to the QIAsymphony SP/AS
instruments. See Section 8.3.3 of the
QIAsymphony SP/AS
User Manual — General Description
for more information.
Check whether the Assay Parameter Set name is spelled
correctly in the work list file. If required, modify the name of
the Assay Parameter Set in the work list and then transfer the
modified work list file to the QIAsymphony SP/AS instruments.
Restart the assay definition process.
75503
Your current process configuration
does not allow you to clear an
assignment of an X to a sample
which is linked to this Y in the work
list.
This sample is linked to the Assay Control Set/Assay
Parameter Set in the work list.
Either use the assignments in the work list, or the “Supervisor”
can set the parameter
Allow information for single samples in
work lists to be overwritten?
to
Yes
in the
General Process
tab
of the
Configuration
menu.
75505
Your current process configuration
does not allow you to assign an X to
a sample which is not linked to this Y
in the work list.
The sample is not linked to the Assay Control Set/Assay
Parameter Set in the work list.
Either use the assignments in the work list, or set the parameter
Allow information for single samples in work lists to be
overwritten?
to
Yes
in the
Configuration
menu.
76202
An unknown kit bar code has been
entered.
Ensure that the entered bar code is either a QIAGEN bar code
or that it meets the requirements for a custom bar code, as
described in “Defining customized kit bar codes” in
Section 2.5.1 of the
QIAsymphony SP/AS User Manual —
Operating the QIAsymphony AS
.
76203
The kit bar code has not been
entered for every assay. Do you want
to continue?
Only one kit bar code for a multi-assay run has been entered
or scanned.
If this was on purpose because the other assays do not have
kit bar codes, press
Yes
to continue.
If this was not on purpose and the other assays do have kit
bar codes, return to the
Loading reagents
screen list view and
enter or scan the kit bar code(s) for the other assays.