QIAsymphony SP/AS Consolidated Operating Guide 12/2017
77
Locations of bar code reader and spacer.
3.
As soon as the bar code reader is in position, the slot unlocks and the green LED starts to
flash. Slide the carrier into the slot until it locks.
4.
The bar code reader reads bar codes on the carrier, inserts, and corresponding sample tubes
(if bar coded). Upon successful loading, the LED changes from green to orange.
5.
The bar code reader returns to the home position.
6.
To add more sample tubes in different slots, follow procedure as described in this section.
Otherwise close the “Sample” drawer.
Important
Be sure to support the tube carrier with your second hand during the loading
process. Otherwise, there is a risk of handle breakage.
Important
Make sure to slide the carrier smoothly into the slot otherwise an error may
occur.
Important
Only a tube carrier containing internal control can be loaded into “Slot A”.
Tube carriers containing samples must be loaded into “Slot 1”, “Slot 2”,
“Slot 3”, or “Slot 4”.
Important
If you need to load two samples with the same bar code/ID in the same tube
carrier, do not place them side by side. Otherwise an error will occur.
Bar code reader (laser)
Spacer