QIAsymphony SP/AS Consolidated Operating Guide 12/2017
113
Important
Be aware that if you use the functions
Modify Run
and
Create AS Batch
after an
integrated run has been queued, the order in which SP and AS batches are
processed by the system may be different to the order in which batches would
be processed if AS batches were created before queuing the integrated run.
For more detailed information, refer to Appendix A of the
QIAsymphony SP/AS User Manual —
Operating the QIAsymphony AS
.
Important
At least one assay has to be assigned to samples of the QIAsymphony SP batch.
Important
This action can also be performed for completed QIAsymphony SP batches,
allowing automated assay setup of samples for which purification has already
been completed.
12.3.2
Loading an integrated run
First, load the QIAsymphony SP. Then load the QIAsymphony AS. This section outlines how to
load samples, reagents, and consumables onto the QIAsymphony AS.
In addition, the
Loading Information
screen provides an overview of which labware,
consumables, and adapters are required for a run. The number and type of filter-tips that are
required is displayed. Press a particular slot for more detailed information.