QIAsymphony SP/AS Consolidated Operating Guide 12/2017
46
6.5.1
Synchronizing files on instruments with files on the USB stick
To synchronize files on QIAsymphony SP/AS instruments with files on the USB stick, follow the
steps below.
1.
Log in to the QIAsymphony SP/AS instruments.
Prepare the USB stick with the files for synchronization. Store the files you want to upload to
the QIAsymphony SP/AS instruments in their corresponding folders on the USB stick (e.g., a
newly defined rack file in the folder
/data/Worklists/
).
2.
Insert the USB stick into one of the USB ports at the front of the QIAsymphony SP.
3.
Press
File Transfer
in the
Tools
screen to enter the
In-/Output Files
menu.
4.
Select one of the file transfer tabs (
In-/Output Files
,
Process Files, Instr. Setup Files
).
For example, to synchronize work lists, select the
In-/Output Files
tab.
5.
Select the file type(s) on the QIAsymphony SP/AS instruments that should be synchronized
with the files on the USB stick by pressing the appropriate button(s) in the
Transfer from USB
stick
panel.
6.
Set
Synchronize files
to
Yes
by pressing the
Yes
button.
7.
Press the
Transfer
button in the command bar of the screen to synchronize the selected files
type(s).
A message appears informing you that the files will be synchronized. Check that the
information is correct.
8.
To continue with the synchronization, press
Yes
.
After successful synchronization, a message will appear confirming synchronization.
9.
Press
OK
to continue.
10.
Remove the USB stick.
6.5.2
Synchronizing files on the USB stick with files on instruments
Files on the USB stick can be synchronized with files on the QIAsymphony SP/AS.
This means that files stored on the QIAsymphony SP/AS are transferred to the USB stick.
If the file already exists on the USB stick it will be overwritten by the file from the
QIAsymphony SP/AS instruments.
Files that exist on the USB stick but do not exist on the QIAsymphony SP/AS instruments are