QIAsymphony SP/AS Consolidated Operating Guide 12/2017
102
12.2
Preparing a run
Before defining a run, available adapter(s) and holder(s) must be configured in the software. If
work list(s) and rack file(s) will be used in the run, these files must be transferred to QIAsymphony
SP/AS instruments.
For detailed information about transferring process files, work lists, rack files, and concentrations
data files, see Section 6.
12.2.1
Assay favorites
In the integrated mode, personalized assay favorites can be defined for quicker assignment within
the setup screen for
Integrated Operation
.
1.
Press the
Tools
tab and select
Assay Favorites
.
The
Define Assay Favorites
screen appears as shown below.
The dialog contains a list showing the available assays for “Integrated Setup” and the set of
favorite buttons identical to the ones in the “Integrated Setup” screen.
2.
Select the assay to assign as a favorite.
3.
Assign the selected assay to a selected blank
Favorites
button.
The assay will be displayed on the assigned favorite button.