QIAsymphony SP/AS Consolidated Operating Guide 12/2017
150
7.
Unload the normalization rack from the drawer.
8.
Press
Remove
in the touchscreen.
9.
Remove empty tip racks.
10.
Empty the tip disposal bag.
11.
Close the drawers and press
Scan
to perform an inventory scan.
When the inventory scan is complete, the
Assay Setup
Overview
screen appears.
12.
Press
Remove
in the assay setup
Overview
screen.
Important
A normalization rack containing unused positions cannot be used for subsequent
runs.
12.4.12
Pausing, resuming, and stopping an independent run
1.
To pause or stop a run when a run is in progress, press
Pause AS
in the
Assay Setup
Overview
screen
2.
After pressing
Pause AS
, the
Continue AS
and
Stop AS
buttons appear. The
run can now be resumed or stopped.
Samples will always be flagged as “unclear” if the run has been paused.
The QIAsymphony AS will complete the current pipetting step before
pausing.
Summary of Contents for QIAsymphony SP
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...