QIAsymphony SP/AS Consolidated Operating Guide 12/2017
192
15
Technical Data
QIAGEN reserves the right to change specifications at any time.
15.1
Environmental conditions
Operating conditions
Power consumption
QIAsymphony SP
100–240 V AC, 50/60 Hz, 800 VA
Power consumption
QIAsymphony AS
100–240 V AC, 50/60 Hz, 600 VA
Mains supply voltage fluctuations are not to exceed 10% of nominal supply
voltages. The inlet is on the QIAsymphony SP; in combined operation, the
maximum power consumption is 1400 VA.
Overvoltage category
II
Air temperature
15–32ºC (59–89.6ºF)
Relative humidity
15–75% (noncondensing)
Maximum 75% relative humidity for temperatures up to 31ºC (88ºF), decreasing
linearly to 50% humidity at 32°C (89.6°F)
Altitude
Up to 2000 m (6500 ft.)
Place of operation
For indoor use only
Pollution level
2
Environmental class
3K2 (IEC 60721-3-3)
3M2 (IEC 60721-3-3)
Transportation conditions
Air temperature
–25ºC to 70ºC (–13ºF to 158ºF) in manufacturer’s package
Relative humidity
Maximum of 75% (noncondensing)
Environmental class
2K2 (IEC 60721-3-2)
2M2 (IEC 60721-3-2)
Storage conditions
Air temperature
5ºC to 40ºC (41ºF to 104ºF) in manufacturer’s package
Relative humidity
Maximum of 85% (noncondensing)
Environmental class
1K2 (IEC 60721-3-1)
1M2 (IEC 60721-3-1)
Summary of Contents for QIAsymphony SP
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...