To stop the UV irradiation procedure before the defined period of time has
elapsed, press
Cancel
. The procedure will stop as soon as the robotic arm
completes the current movement.
Important
UV irradiation of the QIAsymphony SP and AS must be performed sequentially
and cannot be performed in parallel. The second irradiation may be left to run
overnight.
14.7
Maintenance of the tip adapter O-ring
This section describes replacing the tip adapter O-ring using the O-Ring Change Tool Set (cat. no.
9019164) to perform O-ring change. The O-rings must be changed every month using the O-Ring
Change Tool Set.
Before removing the old O-ring, the new O-ring must be prepared. These steps should be
performed for both the QIAsymphony SP and the QIAsymphony AS instruments.
For instructions, refer to the quick guide that is equipped with the O-Ring Change Tool Set. If there
is no O-Ring Change Tool Set available, contact QIAGEN Technical Services