QIAsymphony SP/AS Consolidated Operating Guide 12/2017
116
Loading information
screen with assay racks assigned to slot 5 and 6.
“Assay” slots are assigned automatically by the software, the assignment cannot be changed by
the user. The assignment depends on the processing workflow. Slot 5 is processed first, then
slot 6, and then finally slot 4.
Assigning assay racks
1.
Open the
Assays
drawer. Temporary cooling for the defined slots
is started.
2.
In the
Assay Setup/Loading information
screen, press the first
“Assay” slot to be loaded (highlighted yellow). Detailed loading,
information for the slot is shown
3.
Assign rack type and rack ID.
For details, see this section “Assigning assay racks” or the
following section “Assigning assay rack types”.
4.
Place the empty assay rack in the appropriate adapter on the
correct “Assay” slot(s).
Ensure that the appropriate adapter is used with each assay rack.
5.
Press “Load”. The “Assay Setup/Loading information” screen
appears again. The loaded slot is now blue.
6.
If more assay racks have to be loaded, repeat steps 2–5 for the second assay slot.
7.
Leave the “Assays” drawer open to enable loading of normalization rack (optional) and
disposable filter-tips.