Editing the Current Method
Using Method Mode
4 - 11
4
To move to the TE1 data field, press the
key. The timed event may be either O,
C, or M. These letters indicate the following:
O: Open - The relay opens at the time of the event.
C: Close - The relay closes at the time of the event.
M: Momentary - The relay closes for about two seconds, then opens. If the previous
state of the relay is Close, this action may not cause the desired operation. Momentary
closures should be preceded with an Open event.
enter
5
To add another event, press either the
key or the
key. A second event
will be added and the Time field for the next event will increment by 0.01 minute. You
can add up to 20 events using this process.
enter
insert
TIMED EVENTS SCREEN
6
To save your entries, press the
key. This also takes you back to the Method
Screen and saves the changes until you turn the power off on the Detector. When you
restore power, the default values for the method will also be restored. To permanently
store the changes, refer to the “Storing a Method” section in this chapter.
return
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...