Safety and Regulatory Information
Safety and Regulatory Information
S - 16
Solvents
•
Always use clean solvents. Solvents which have been distilled in glass (HPLC Grade) are
recommended.
•
Filter the solvents and buffers through a 0.5-micron medium as an additional precau-
tion.
•
Degas all aqueous and most organic solvents prior to use.
•
Store flammable solvents or solvents which may form hazardous by-products when the
instrument is shut down, by following the recommended shutdown procedure.
•
Check compatibility of solvent(s) with the type of column(s) being used.
•
Know the relative polarity and miscibility of the solvents being used.
Solvents with Low Boiling Points
Do not use liquids that have a boiling point less than 30 °C.
Buffers
Exercise care when using buffers in conjunction with organic solvents. NEVER LEAVE
BUFFERS IN THE SYSTEM OVERNIGHT. Buffers left in the system can form salt crystals
which may cause premature injector valve failure and plug the transfer tubing and sample
needle. To remove buffers, flush the system with water followed by methanol or isopropanol.
Remember to change the flush solvent from methanol or isopropanol to water before using
buffers.
Do not use carbon disulfide or other solvents
which have an auto-ignition temperature
below 110 °C.
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...