![PerkinElmer 200 UV/VIS Series User Manual Download Page 142](http://html1.mh-extra.com/html/perkinelmer/200-uv-vis-series/200-uv-vis-series_user-manual_1530268142.webp)
Maintenance
Routine Maintenance
7 - 12
5. Unscrew the knurled screw that is holding the front panel in place.
6. Swing the front panel open to the right.
7. Slide the drawer out to the first stop.
8. Loosen the Allen bolt on the lower left side of the drawer with the special Allen wrench
provided. (See picture below.)
9. Slide the drawer completely out of the cover.
10. Unfasten the lamp retainer spring.
11. Grip the entire lamp with one hand while holding the detector steady with the other.
Slide the entire lamp assembly off the lamp mount.
12. Unplug the lamp.
Figure 7-5.
Preparing to slide the drawer out of the cover
Avertissement
WARNING Hot
Replace lamp with PerkinElmer part
number N292-0149.
LA LAMPE EST TRÈS CHAUDE!!
Remplacer la lampe par une lampe
PerkinElmer, numéro de pièce N292-0149.
Do not disturb the black fan connector
located below the white lamp connector. If
the fan connector is not properly connected,
the detector will not operate.
Allen bolt
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...