![PerkinElmer 200 UV/VIS Series User Manual Download Page 165](http://html1.mh-extra.com/html/perkinelmer/200-uv-vis-series/200-uv-vis-series_user-manual_1530268165.webp)
Application Notes
Principles of Operation
8 - 9
There are various forms of noise. Electronic noise can arise from line voltage fluctuations,
the detector's circuitry, and radio frequency interference. Short term noise has frequencies
that range from 1 cycle/1 sec to 1 cycle/5 min. The very wide peaks and valleys observed in
long term noise can arise from system variations such as flow rate and pressure variations,
teaks, temperature variations, aging lamps, and gas in the detector cell.
Response Time Versus Time Constant
The time constant value describes the degree of filtering of frequencies. For a single pole fil-
ter, the time constant (
τ
) is the time it takes the output signal to reach 63.2% of the value of
an input signal change. In 2
τ
, 86.5% of the final value is reached, and 95% is reached in 3
τ
. It
is important to realize that time constant is a valid mathematical description for only single
pole filters.
Figure 8-4.
Definition of rise time on a response versus time curve.
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...