Setting Configuration Options
Configuring the Detector
6 - 17
Press F3
to move to the Wavelength Calibration Check Screen. The screen
instructs you to remove the flowcell from the Detector and replace it with a Holmium
Oxide Filter. For information on the procedure for removing and replacing the flow-
cell, see Chapter 7, Maintenance.
λ
CHK
2
Press the
softkey to run the wavelength calibration check. The Detector
checks for the Holmium Oxide Filter and then runs the calibration test.
STRT
3
When the test is complete, the Detector displays the results of the test and instructs
you to remove the Holmium Oxide Filter. Press the
key to return to the Con-
figure Screen.
If the wavelength calibration consistently produces incorrect values, contact your
PerkinElmer service representative. The detector can only be recalibrated by a
PerkinElmer engineer. Your service representative can perform monthly checks on
the wavelength calibration if your lab requires it.
return
4
WAVELENGTH CALIBRATION CHECK SCREEN
WAVELENGTH CALIBRATION - RESULTS SCREEN
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...