![PerkinElmer 200 UV/VIS Series User Manual Download Page 122](http://html1.mh-extra.com/html/perkinelmer/200-uv-vis-series/200-uv-vis-series_user-manual_1530268122.webp)
Configuring the Detector
Setting Configuration Options
6 - 12
Adjusting the “Smoothing” Mode
The “Smoothing” Mode options allow you to reduce noise coming from the detector. There are
two available options, RSP and PW. The two options provide different ways to set the
parameters; however, both allow you to “smooth out,” or filter the peaks in the data to reduce
noise. When you press the SMTH softkey from the Configure menu, the Detector displays
the Smoothing Mode Screen. This screen allows you select one of the following:
•
RSP - Response Time in Seconds
Indicates how fast the slope of a peak changes.
•
PW - Peak Width at Half Height in Seconds
Allows you to set the smoothing parameter in terms of how far the peak rises.
For information on selecting RSP or PW as a control parameter, see Chapter 3, Using Man-
ual Mode or Chapter 4 Using Method Mode. For information on how to select the correct
value for the RSP or PW, see Chapter 8, Principles.
Press the corresponding softkey to select the
or
option.
PW
RSP
SMOOTHING MODE SCREEN
From the Instrument Setup Screen, press F5
to move to the Smoothing Mode Screen.
SMTH
1
2
3
To accept and save the change(s) and return to the Instrument Setup Screen, press
the
key.
To cancel changes and return to the Instrument Setup Screen press the
key.
The changes you make on this screen are reflected on the Method Screen, Default Manual
Settings Screen, and Scan Screen.
return
quit
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...