Making Input/Output Connections
Installation
2 - 11
Making Input/Output Connections
The input/output connections on the Series 200 UV/VIS LC Detector uses phone jack connec-
tors for quick contact closure connections between instruments. Figure 2-3 shows the input/
output connectors on the right-side panel of the Series 200 UV/VIS LC Detector. Table 2-4
provides a description of each input/output connector.
Three wire terminal blocks at the top of the Connector have removable plugs, each of which
are keyed and can only be inserted with the wires exiting from the topside of the plug.
Figure 2-4.
The Series 200 UV/VIS LC Detector Terminal Connectors.
Table 2-4.
Input/Output Connectors.
Input\Output Connector
Description
EXT RUN
Actuating this input from an external contact closure starts
the method. This performs the same action as the Run key.
You can run a method when READY appears on the top line
of the method display.
EXT STOP
Actuating this input from an external contact closure stops
the detector. This performs the same action as the Stop key.
RUN OUT
This contact closure output signals the start of a run. This
contact switches for approximately two seconds at the
instant a method is started (at the moment Step 1 starts).
RDY
This contact closure output signals that initial conditions
have been reached and a method is ready to run. When
READY appears on the top line of the method, this contact is
switched and held until the method is run.
TE1
This contact closure actuates at a designated time in the
method and controls an external device. When activated, the
TE1 contacts are switched and held for two seconds, if set to
momentary, or set to opened or closed.
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...